Programming And Calibrating The Area Dose Product Measuring Device; Programming - Siemens ARCADIS Avantic Installation And Start-Up Manual

Option area dose measurement
Hide thumbs Also See for ARCADIS Avantic:
Table of Contents

Advertisement

6
Programming and calibrating the area dose product measuring
device
NOTE

Programming

Start and log on to the service program.
<Click Main System> and <Next>.
Click <Options> under "Configuration".
Click <Yes> under "Diamentor Existing".
Read off the "TW" factor from the identification tag on the diamentor chamber and enter
it in the "Constant" field.
Select ""0.1 cGy cm*cm/pulse" for the standard resolution in the "Resolution" field
(see
For pediatric applications select
"0.01 cGy cm*cm/pulse" under "Resolution".
Enter the test value (TW) 500 under "Constant" (see
5000 for pediatrics.
Press the Return key on the keyboard and click <Save>.
ARCADIS Avantic
Installation and Setting Instructions
This calibration operation is used to adjust the area dose product
measuring device to the ARCADIS Avantic system display.
(1/Fig. 2 / p.
7)).
SPR2-330.814.01.01.02
02.05
(2/Fig. 2 / p.
7)) for standard and
Page 6 of 10