8

Final work steps

Click <Backup & Restore> in the "Home menu", select "Backup" under "Command",
and select "(-R-) CD-R" under "Drives".
Use <Go> to start the burning operation for the following packages: SW-Settings.., As-
piaSettings, Security-Settings, ExamSet, MainSystem.
Perform partial acceptance tests in accordance with country-specific regulations.
Check the calibration of the laser light localizer if applicable.
ARCADIS Avantic
Installation and Setting Instructions
SPR2-330.814.01.01.02
02.05
Page 8 of 10
CS SD 24
0
Siemens AG
Medical Solutions