Programming the module
Depending on the data system, program the module using these data system
screens:
•
In MassLynx, the Inlet Method Editor (see the MassLynx Help).
•
In Empower, the Instrument Method Editor dialog box (see the
Empower Help).
•
In Millennium
Control Module Instrument dialog box (see the Millennium
Maintenance and troubleshooting
Warning:
remove the module's protective panels behind which are no
user-replaceable components.
Cleaning the cabinet
Use a soft cloth dampened with water to clean the module's surface if it is
dirty.
Replacing the fuse
To replace the fuse
1.
Power-off the module, and then disconnect the power cord.
2.
Use a small screwdriver to remove the fuse holder from its socket in the
rear panel.
3.
Remove the faulty fuse from the fuse holder.
4.
Install a new, F 3.15-A fuse in the fuse holder (see
5.
Refit the fuse holder into its socket.
2-14
Installing the Pump Control Module II
32
Chromatography Manager 3.2 and 4.0, the Pump
To prevent injury to yourself or damage to the module, do not
32
Help).
Appendix
C).