Left terminal strip positions (Continued)
Left connector
SW 3 (two)
GND
INJECT START
GND
Right terminal strip positions
Right connector
SW 4 (two)
SW 5 (two)
SW 6 (two)
GND
STOP FLOW
GND
Event cables
Caution:
To avoid damaging peripheral devices, power them off before
connecting to or disconnecting cables from the module.
To connect the event cables
1.
Remove a terminal strip from the module by grabbing each end and
carefully pulling it from the module.
2.
Connect the two spade connectors at the two-wire end of a shielded
event cable to one of these:
A manual injector, for an event in signal
A device to be triggered, for an event out signal
3.
Cut off the spade connectors from the red and black leads on the
three-wire end, and then strip 1.5 cm from the ends.
Description
Relay 3
Active Low
Description
Relay 4
Relay 5
Relay 6
Active Low
I/O type
Out
Ground
Input
Ground
I/O type
Out
Out
Out
Ground
Input
Ground
Rear-panel signal connections