Physical And Environmental Specifications; Accuracy And Precision - Waters Pump Control Module II Installation Manual

Table of Contents

Advertisement

Physical and environmental specifications

Physical and environmental specifications
Parameter
Dimensions
Weight
Operating
temperature
Operating humidity 10 to 90%, noncondensing

Accuracy and precision

Pump pressure specifications
Parameter
Input range
Pressure update
rate
Note:
Flow precision and accuracy is determined by the type of pumps that
the module controls (Waters 510 or 515 HPLC pumps).
B-2
Specifications
Allowed values
14 cm wide × 41 cm deep × 20 cm high
Approximately 2.7 kg
4 to 40 °C (39 to 104 °F)
Allowed values
−10 to 0 VDC, representing a pressure range from 0 to
41365 kPa (413.7 bar, 6000 psi)
10 kHz