5
Maintenance
5.1
Cleaning and disinfecting
NOTICE
Danger of material damage!
Do not use solvents (e.g., acetone, alcohol) or cleaning
agents with abrasive particles for cleaning and disinfecting.
► Only apply the described procedures and use the cleaning
agents and disinfectants specified for powered air-
purifying respirators.
5.1.1 Manual cleaning
For information on suitable cleaning agents and disin-
fectants and their specifications, see document
9100081 at www.draeger.com/IFU.
Dräger recommends MediClean Forte or Suma® Light
D1.2 for the manual cleaning of the X-plore 8000 face shield
and X-plore 8000 helmet with PC visor HL1.
1. Prepare a cleaning solution of water and the cleaning
agent. Alternatively, you can use lukewarm soapy water.
2. Clean facepiece in an immersion bath.
3. Rinse with fresh water.
4. Allow all parts to air-dry or leave them to dry in a drying
chamber (temperature: max. 50 °C, time: max. 2 h).
Protect from direct sunlight.
5. Check all components. Replace damaged or deformed
components, or components that are stuck together.

5.1.2 Mechanical cleaning

Only clean and disinfect face cuff manually. Mechanical
cleaning and disinfection are not recommended for this part.
High temperatures during cleaning and disinfection put a lot of
strain on the materials. Dräger recommends using the lowest
possible temperatures (max. 55 °C). The facepiece is not
suitable for washing machines and should therefore only be
spray-cleaned.
1. Remove face cuff.
2. Load a cleaning/disinfection unit with the facepieces.
3. Start the standard mask program (spray program) with the
following settings:
Duration of cleaning phase: 5...20 min.
Water temperature: 55 °C
Cleaning agent: 0.4% Suma Jade Pur-Eco L8
Rinsing phase: 1x with 0.05% Suma Med Neutral
4. Remove facepieces from the machine.
5. If the facepiece is still dirty, clean it again.
6. If necessary, remove detergent residues on hard plastic
parts with a cloth and clean water.
7. Allow all parts to air-dry or leave them to dry in a drying
chamber (temperature: max. 50 °C, time: max. 2 h).
Protect from direct sunlight.
8. Check all components. Replace damaged or deformed
components, or components that are stuck together.
Instructions for use
|
X-plore 8000 Face Shield X-plore 8000 Helmet PC visor HL1
9. Mount face cuff.
5.2

Maintenance work

For information on spare parts, see
https.://www.connect.draeger.com.
5.2.1 Replacing the protective foil
1. Remove the old protective foil and adhesive strips, if still
attached.
2. If necessary, wipe the visor shield with a damp cloth and
dry with a soft cloth.
3. Center the new protective foil on the visor frame and stick
it to the visor shield.
5.2.2 Changing the visor shield
1. Pull up the cross-bars and remove the visor fasteners on
both sides of the visor (see fold-out page, figures D1, D2).
Store each visor fastener with its associated cross-bar.
2. Push visor shield out of the visor frame at the upper edge
from within (see fold-out page, figures D3, D4).
3. Insert a new visor shield from outside into the slot of the
visor frame (see fold-out page, figure E1). Start on one
side to do this. Next, push the bottom of the visor shield
into the slot. Next, push the other side of the visor shield
into the slot (see fold-out page, figure E2, E3).
4. Push the visor shield into the three recesses of the visor
frame at the top, until they audibly click into place (see
fold-out page, figure E4).
5. Insert visor fasteners into the visor frame on both sides.
Insert the cross-bars into the visor fasteners and push
them firmly downwards until they audibly click into place
(see fold-out page, figure E5).
5.2.3 Changing the visor frame
1. Push visor frame fastener out of the visor frame with a
suitable tool (see fold-out page, figure G1).
2. Remove matching parts from visor frame (see fold-out
page, figure G2).
3. Remove visor frame from facepiece (see fold-out page,
figure G3).
4. Place a new visor frame on the bump cap/helmet shell and
fasten with visor frame fasteners and matching parts.
The visor fasteners are labeled with "R" and "L". Make
sure that they are inserted in the right side.
5. If necessary, insert a new visor shield into the visor frame.
5.2.4 Changing the padding
1. Pull the padding from the hook-and-loop points of the
support ring.
2. Center and attach the new padding.
|