14.1 Sensor diagnostics

Sensor diagnostics
Calibration timer
Service timer
Sensor

14.2 Language selection

Language selection
Check the language as required and restart the instrument to apply the change. The instrument will
restart in the language selected.

14.3 Clock

Clock
Type in each appropriate box the actual time and date, and select the display format for them.
The instrument can automatically remind the user when the next sensor
calibration is due.
Select enable and enter a delay in days.
The display shows the current instrument date and time, next calibration
due date and time, and the remaining days.
The next calibration date is updated when the sensor is calibrated. The
event "Cal. required" is generated when the delay has elapsed.
The instrument can automatically remind the user when the next sensor
service is due.
Select enable and enter a delay in days. This should be validated by a
level 3 user.
The display shows the current instrument date and time, the next sensor
service due date and time, and the remaining days.
The next service date is updated when the button Service done is pressed
after a service. The event "Service required" is generated when the delay
has elapsed.
The sensor attached to your instrument will require periodic servicing and
maintenance. For more information on this, please refer to the manual
delivered with the sensor.
Select either General or Zero diagnostic from the menu.
General will display values for the measurements in the sample, while Zero
is used to display values when the gas concentration is at zero (e.g. during
a zero calibration using pure nitrogen as the calibration medium).
The screen layouts are identical for both diagnostic options, and show
useful information for troubleshooting purposes.
The first screen displays measurement information, with a status (OK or
NOK) alongside. A status of NOK indicates the value is outside of the
acceptable minimum or maximum value. To view the acceptable range limits
press the Show criteria button.
The criteria screen displays the acceptable range limits (minimum and
maximum values) for more in-depth troubleshooting.
Note: If Zero diagnostic is selected, a warning message appears. Ensure
the sensor has been calibrated in pure N
Services Menu
before selecting this option.