Hardware Description - Hach ORBISPHERE K-M1100 User Manual

Analyzer
Hide thumbs Also See for ORBISPHERE K-M1100:
Table of Contents

Advertisement

Specifications
The oxygen partial pressure (pO
) is then linked to the corresponding phase-shift measurement
2
( to build the sensor calibration curve (shown right in
Figure
3). This curve is described by the
Stern-Volmer equation (shown left in
Figure
3) where K
is the indicator quenching constant (in
sv
-1
mbar
) representing the quenching efficiency of the oxygen and thus the sensor sensitivity, f
is
0
a constant and 
is the phase-shift at zero oxygen representing the unquenched fluorescence
0
decay time of the dye. The calibration curve thus relies on two parameters: the phase-shift at
zero oxygen and the luminescent spot sensitivity, K
. The dissolved oxygen concentration is
sv
then calculated with Henry's law using the water solubility curve as a function of the
temperature.
Figure 3 Stern-Volmer equation and calibration curve

2.4 Hardware description

The instrument hardware is made of one main board, and one measurement board for the
measurement channel (= the sensor).
The main board includes the controls for power, display, the touch screen, the barometric
sensor, the alarms, and communication ports. The measurement board performs measurements
and executes commands from the main board. It holds the "Analog output" and "Relays" that
send information to external systems.
A hardware watchdog is activated at program start up, to check that the system is not frozen
(i.e. infinite loop, system crash, etc.). If the watchdog is not refreshed by the software every
minute, the measurement display, the relays and the analog output are frozen up to 2 minutes.
Then the reset shuts down the instrument for 10 seconds and the start-up procedure is
performed. At the same time all the hardware (sensor, measurement board) are reset.