Installation

3.3 Connecting sensor cable

Fig. 4
Connection of the sensor plug to the controller
Fig. 5
Sensor connector pin assignment
1
6
5
4
Number
1
2
3
4
5
6
14
CAUTION
Always lay cables and hoses so that they do not pose a tripping risk.
1. Unscrew the protective caps from the controller socket and the cable plug and
retain them.
2. Pay attention to the guide in the plug and slide the plug into the socket.
3. Tighten the nut by hand.
Note: Extension cables are available in various lengths (refer to
parts and accessories, page
Maximum cable length 100 m (328 ft).
2
3
Description
+12 VDC
Ground
Data (+)
Data (–)
Screen
Notch
29).
Section 7, Replacement
Cable colour
brown
black
blue
white
Screen (grey)