Sensor Diagnostics Menu; Sensor Setup Menu - Hach 5740sc User Manual

Galvanic membrane dissolved oxygen sensor
Table of Contents

Advertisement

4.3 Sensor Diagnostics Menu

SELECT SENSOR
ERROR LIST–See
section 7.1 on page
WARNING LIST–See
section 7.2 on page

4.4 Sensor Setup Menu

SELECT SENSOR (if more than one sensor is attached)
CALIBRATE
AIR CAL
Perform an air calibration of the sensor (slope calibration). See
SAMPLE CAL
Enter a value for the DO concentration as determined by another sensor or independent method. The instrument
performs an offset calibration based on the entered value. See
TEMP ADJUST
Displays the measured temperature and allows the user to adjust the temperature by ±15 °C.
DEFAULT SETUP
Restores the gain and offset values to 1.0 and 0.0, respectively, and restores the sensor code to default.
CONFIGURE
EDIT NAME
Enter up to a 10-digit name in any combination of symbols and alpha or numeric characters.
MEAS UNITS
Select the appropriate measurement units to display. Choose from: mg/L, ppm, or percent. Default: ppm
TEMP UNITS
Select Celsius (°C) or Fahrenheit (°F); Default: °C
AC FREQUENCY
Choose 50 or 60 Hz depending on the power line frequency for optimal noise rejection. Default is 60 Hz.
FILTER
Specify the number of seconds for signal averaging (0–60). Default: 0 seconds.
PRESSURE UNITS
Choose pressure units in mmHg, feet, meters. Default: mmHg
SET PRESSURE
Enter either altitude or air pressure. Correlates to the pressure units setting. Range: –5000 to 15000. Default: 760 mmHg
SALINITY UNITS
Choose from mS/cm, mMol/L, ppt, or mg/L. Default: mS/cm
SET SALINITY
User-entered value. Default: 0.00 µS/cm
LOG SETUP
Disable or choose the datalogging interval for sensor and temperature measurements. Use the arrow keys to move
through the available choices for each. Default: Disabled
37.
37.
section 4.5.1 on page
section 4.5.2 on page
sc100 Operation