Hach ORBISPHERE K-M1100 User Manual page 85

Analyzer
Hide thumbs Also See for ORBISPHERE K-M1100:
Table of Contents

Advertisement

Communication Menu
Commands
The "Command Output Buffer" is formatted as follows:
Output command toggle (OCT)
Output command ID (OCI)
Output command data byte 1 (OCD1)
Output command data byte 2 (OCD2)
Output command data byte 3 (OCD3)
Output command data byte 4 (OCD4)
The "Command Input Buffer" is located just after the measurement data and is formatted as
follows:
Input command toggle (ICT)
Input command status (ICS)
Input command data byte 1 (ICD1)
Input command data byte 2 (ICD2)
Input command data byte 3 (ICD3)
Input command data byte 4 (ICD4)
The following commands are available:
Change product
Activate sensor (valid for EC sensors only)
Set channel hold status (valid for TC and LDO sensors only)
Change product command - output
Change product command - input
84
Name
Name
Name
Value
OCT
1-2
OCI
1
Command ID
Channel number:
OCD1
0
0 = Channel 1
OCD2
0-99
Product number
Erase measurement files:
OCD3
0-1
0 = Never erase the measurement files.
1 = Erase measurement file if necessary (e.g. gas unit changes)
OCD4
Not used
Name
Value
ICT
1-2
0 = OK
1 = Unknown command ID
ICS
0-3
2 = Invalid parameter (e.g. invalid channel no or product number)
3 = Execution failure
ICD1
Not used
ICD2
Not used
ICD3
Not used
ICD4
Not used
Type
Output byte
Output byte
Output byte
Output byte
Output byte
Output byte
Type
Input byte
Input byte
Input byte
Input byte
Input byte
Input byte
Comment
Comment
Size
Offset
8 bits
0
8 bits
1
8 bits
2
8 bits
3
8 bits
4
8 bits
5
Size
Offset
8 bits
28
8 bits
29
8 bits
30
8 bits
31
8 bits
32
8 bits
33