Logic I/O - Emerson Rosemount Analytical Oxymitter 4000 Instruction Manual

Hazardous area oxygen transmitter
Table of Contents

Advertisement

Instruction Manual
IB-106-340C Rev. 4.1
July 2004
The HART option is not protected by
energy limiting barriers. It must not be
interfaced from within the hazardous
area. The 4-20 mA cables should be
routed and the connections made out-
side the hazardous area. Note that this
is the case even when using the intrin-
sically safe version of the handheld
communicator.
3-2

LOGIC I/O

This two-terminal logic contact can be config-
ured either as a solid-state relay-activated alarm
or as a bi-directional calibration handshake sig-
nal to an IMPS 4000 or SPS 4000. The configu-
ration of this signal depends on the setting of
the LOGIC I/O PIN MODE via HART/AMS or
LOI. The ten different modes available are
explained in Table 3-1.
a. Alarm
When configured as an alarm, this signal
alerts you to an out-of-spec condition. The
output is +5 Vdc in series with a 340 ohm
resistor.
Table 3-1. Logic I/O Configuration (as set at HART/AMS or LOI)
Mode
0
1
2
3
4
5*
6
7
8**
9
* The default condition for a Hazardous Area Oxymitter 4000 without an IMPS 4000 or SPS 4000.
** The default condition for a Hazardous Area Oxymitter 4000 with an IMPS 4000 or SPS 4000.
3-4
Configuration with Keypad
The unit is not configured for any alarm condition.
The unit is configured for a Unit Alarm.
The unit is configured for Low O
The unit is configured for both a Unit Alarm and Low O
The unit is configured for a High AC Impedance/CALIBRATION
RECOMMENDED.
The unit is configured for both a Unit Alarm and a High AC Imped-
ance/CALIBRATION RECOMMENDED.
The unit is configured for both a Low O
ance/CALIBRATION RECOMMENDED.
The unit is configured for a Unit Alarm, a Low O
pedance/CALIBRATION RECOMMENDED.
The unit is configured for a calibration handshake with IMPS 4000 or
SPS 4000. CALIBRATION RECOMMENDED will initiate the calibra-
tion cycle.
The unit is configured for a calibration handshake. CALIBRATION
RECOMMENDED will not initiate the calibration cycle with the IMPS
4000 or SPS 4000.
Hazardous Area Oxymitter 4000
For optimum performance, Rosemount rec-
ommends connecting the output to a Potter
& Brumfield 3.2 mA DC relay (P/N R10S-
E1Y1-J1.0K).
Of the ten modes in Table 3-1, modes 0
through 7 are the alarm modes. The factory
default is mode 5 for Hazardous Area Oxy-
mitter 4000 units without an IMPS 4000 or
SPS 4000. In this mode, the output will sig-
nal when a unit alarm or a CALIBRATION
RECOMMENDED indication occurs.
b. Calibration Handshake Signal
If using an optional IMPS 4000 or SPS
4000, the logic I/O must be configured for
calibration handshaking. Of the ten modes
in Table 3-1, only modes 8 and 9 are con-
figured for calibration handshaking. For a
Hazardous Area Oxymitter 4000 with an
IMPS 4000 or an SPS 4000, the factory
sets the default to mode 8. In this mode,
the logic I/O will be used to communicate
between the Hazardous Area Oxymitter
4000 and the sequencer and to signal the
sequencer when a CALIBRATION REC-
OMMENDED indication occurs.
Configuration
.
2
and High AC Imped-
2
, and a High AC Im-
2
Rosemount Analytical Inc. A Division of Emerson Process Management
.
2