Upon completing installation, make sure that the Hazardous Area Oxymitter
4000 is turned on and operating prior to firing up the combustion process.
Damage can result from having a cold Hazardous Area Oxymitter 4000 ex-
posed to the process gases.
During outages, and if possible, leave all Hazardous Area Oxymitter 4000
units running to prevent condensation and premature aging from thermal
cycling.
If the ducts will be washed down during outage, MAKE SURE to power down
the Hazardous Area Oxymitter 4000 units and remove them from the wash
area.