For Your Safety and that of Your Patients For Your Safety and that of Your Patients Strictly follow the Instructions for Use Safety instructions Any use of the apparatus requires full understanding and strict observation of these instructions. The apparatus is only to be used for purposes specified here. It is the responsibility of the doctor to draw conclusions from the skin temperature measurements taken.
Intended Use Intended Use Babytherm 8004/8010 is an open care unit for warming premature babies, neonates and infants with a body weight of up to 8 kg. The unit may be used in operating theatres, neonatal wards, children’s wards, obstetric units and paediatric intensive care units.
Page 5
Intended Use Babytherm 8004 Babytherm 8010 Babytherm 8004 Babytherm 8010 °C °C °C Check Check > 38°C Inop. Inop. < 36°C skin °C skin °C matress °C manual manual skin skin man. man.
Operating Concept Operating Concept Babytherm 8010 °C °C Check > 38°C Inop. < 36°C skin °C manual mattress.°C skin man. ON/OFF indicator: Display and keypad for the radiant warmer: The green LED is lit when the unit is ON. top: display for the measured value of the core and The red LED is lit following a power failure.
Preparation Preparation Shelf The unit is fitted with one shelf. Make sure that the shelf is firmly in position. To fit rails to the shelf: 1 Screw rail to the left and/or right of the shelf. Place desired auxiliary equipment on the rails and shelf. Max.
Preparation X-ray tray Babytherm 8004 only Do not use the X-ray tray in the extended position to write on. Do not place objects other than x-ray plate. Do not lean on the tray. Risk of damage. Preparation 1 Push X-ray tray fully home in the horizontal position.
Preparation Warming bed The bed is enclosed by two side panels and two end panels that can be opened independently of one another. The side panels are provided with holes for secretion and drainage hoses. The end panels have flexible grommets for hoses and cables.
Preparation Tilting the bed 1 Pull the handle of the locking mechanism out towards the front. 2 Press handle down = head-up position Pull handle up = head-down position Release the handle: the bed automatically locks into the selected position. The bed can be tilted in finely graduated steps.
Preparation Bronchial aspirator Fastening the ejector on the compact rail Fit the compact rail to the desired position on the stand, see page 7. 1 Clip the ejector to the compact rail. 2 Mount bracket with projecting support to one of the positions provided on the column.
Preparation Accessories for oxygen therapy flowmeter Prepare the flowmeter in accordance with its specific Instructions for Use. To mount the flowmeter on the compact rail: 1 Press the slider all the way down; hook the flowmeter to the compact rail and release the slider. 2 Screw on the O connecting hose.
Preparation Fitting the hose bracket Insert the hose bracket into the hole in the left side of the head panel and tighten with knurled screw. Drainage canister hook 1 Insert hook horizontally into hole in the Babytherm casing. 2 Pivot hook downwards. Compressed air distributor / socket strip Secure the compressed air distributor to the standard rail.
Testing Readiness for Operation Testing Readiness for Operation Before using for the first time 1 Check that the mains voltage matches the values indicated on the rating plate next to the power cable. 2 Plug into the mains. Before each use Check that the unit has been disinfected.
Page 15
Testing Readiness for Operation Switching on and activating the self-test Press the On/Off button until it engages = ON. Babytherm now runs a self-test to check important functions. 1 The following message appears in the display: »All displays on, horn on«. Babytherm 8010 All the displays are lit for about 2 seconds: digital displays read 88.8 and a continuous tone sounds.
Page 16
Testing Readiness for Operation Testing the power failure alarm Disconnect the power plug. Babytherm 8010 1 The red LED should light up and a continuous alarm tone should sound. 15min °C Reconnect the power. The unit will continue to operate with the values set before the power failure alarm.
— Keep oxygen fittings and seals free of oil and grease. — Open the valves slowly. Do not use a gel mattress with Babytherm 8004. — Do not use Babytherm in the presence of flammable Do not use a foam mattress with Babytherm 8010.
Page 18
Operation X-ray tray Do not use the X-ray tray in the extended position to write on. Do not place objects other than x-ray plate. Do not lean on the tray. Risk of damage. Remove X-ray cartridge from the X-ray tray for transport. Bed canopy The infant’s core temperature should be monitored when the bed canopy is in use.
Operation Switching on After switching on, the system runs a self-test and then proposes the following default operating mode for the radiant Babytherm 8010 warmer: — if no skin temperature sensors connected = man (manual) 15 min °C — if yellow skin temperature sensor for core temperature is connected = skin (baby control mode) <...
Page 20
Operation 15 minutes alarm To remind the user to monitor the core temperature constantly at higher heater output levels, — an audible alarm is emitted every 14 minutes at heater level 4 and above. At the same time: 1 the yellow LED flashes. —...
Operation Using baby control mode In this operating mode, the skin temperature of the infant is adjusted towards the set value. The sensor attached to the skin measures the skin temperature. The radiant warmer output is adjusted according to the temperature difference between the skin temperature and the desired value.
Page 22
Operation Allow at least 5 minutes for the skin temperature sensor to adjust to the temperature of the infant. If a skin temperature sensor is connected when the apparatus is switched on: 1 Babytherm proposes "baby control mode" as the default operating mode.
Page 23
Operation Outside measuring range If the temperature is outside the measuring range of 15 to 42 3 dashes at the bottom of the display = temperature °C below 15 3 dashes at the top of the display = temperature above 42 Refer to ”Troubleshooting”, see page 46.
Page 24
Operation For deviations greater than ±0.5 C between the set and measured skin temperature: 1 The skin temperature display starts flashing. Babytherm 8010 — An audible alarm sounds, 2 the yellow LED starts flashing and — the red central alarm lamp on the radiant warmer starts to flash.
Page 25
Operation Skin temperature above 39 1 The skin temperature display starts flashing. Babytherm 8010 — The audible alarm sounds. 2 the red LED starts flashing. — The red central alarm lamp on the radiant warmer starts 15min °C to flash. 3 The following message appears in the text display: <...
Operation ThermoMonitoring For improved feedback on the thermal condition of the patient, we recommend that both the core and the peripheral temperature should be measured. Connecting the peripheral skin temperature sensor 1 Plug the white sensor into the white socket. 2 Route the sensor cable through one of the flexible grommets in the bed.
Operation Using heated gel mattress Observe the warm-up time of the gel mattress. Wait until the desired mattress heating value is reached before placing the infant in the Babytherm 8010. Do not switch off the mattress heater while an infant is lying on the gel mattress.
Page 28
Operation Extending lower limit of the setting range C to 30 Only use low temperatures if prescribed by doctor. Monitor patient very closely. 1 Press button until 2 the set value 36 C is displayed. Babytherm 8010 3 The following message appears in the text display: »Mattress temperature Confirm temp.
Page 29
Operation Deviation from set temperature If the deviation between the set and measured mattress temperature is greater than ±1 — The audible alarm sounds, 1 the yellow LED starts flashing and Babytherm 8010 — the red central alarm lamp on the radiant warmer flashes. 2 Measured mattress temperature value flashes, 15min °C...
Page 30
Operation Mattress temperature above 40 1 Red LED flashes — The red central alarm lamp starts to flash. Babytherm 8010 2 Display flashes, 3 The following message appears in the text display: »Mattress temperature above 40 15min °C Press OK button to acknowledge« <...
Operation Using the bed canopy When the bed canopy is in place, the side panels can be folded down. Do not leave the Babytherm unattended with the side panels open, as the infant can fall out. When using external phototherapy, remember the maximum load on the bed canopy is 11 kg.
Page 32
Operation Opening / closing the bed canopy 1 Unlock bed canopy = turn lock 2 Open bed canopy = swivel handles upwards until they engage. The infant is now accessible for care/medical treatment. 2 Close bed canopy = swivel handles forwards. Removing the bed canopy 3 Put bed canopy in closed position = turn lock until it engages.
Operation Recommended heater settings for heat therapy Instructions for setting the mattress heater and radiant warmer, alone or in combination. Using the mattress heater with a gel mattress without using the radiant warmer (Babytherm 8010 only) Normal setting: The core temperature of the patient adapts to the mattress temperature over a relatively long period.
Page 34
Do not cover or dress the patient. Otherwise, the effect of the radiant warmer is uncontrollably reduced. Radiant warmer without mattress heater Babytherm 8004 1 Use foam mattress. Do not use a gel mattress with Babytherm 8004. The unheated gel mattress would cool the patient.
Page 35
Operation Using radiant warmer in combination with mattress heater and gel mattress (Babytherm 8010 only) Normal setting: Set the mattress temperature to the core temperature appropriate for the patient, e.g. between 38.0 C and 38.3 for a premature baby and 37.0 C for a neonate born after the normal term.
Operation Using phototherapy, optional Always use patient eye protection when using phototherapy. Always monitor the differential diagnosis for the neonate to ensure that no life-saving measure is delayed. The patient’s bilirubin levels must be measured regularly. In manual mode, reduce the heat output of the radiant warmer. The phototherapy light directs additional warmth to the patient.
Operation Switching lighting On/Off 1 Press button. The bed surface will now be illuminated by work light. Babytherm 8010 1 Press button again. The work light is switched off. 15min °C 2 Press button. The bed surface will now be illuminated by night light.
Operation Configuration mode No care/therapy should be carried out during configuration of the system, because the normal equipment functions (e.g. temperature measurement, alarms) are switched off. Capabilities available in configuration mode Code Activity Show current software version Set skin temperature deviation Set volume of soft alarm Operating hours of phototherapy Test control panel buttons...
Page 39
Operation Show software version 1 The following message appears in the text display: Babytherm 8010 »Configuration Mode Mode c01: Software version«. 2 The current software version is indicated: e.g. »1.n«. 3 The code for the current mode is displayed: »c01« 15min °C <...
Page 40
Operation Show operating hours for phototherapy 1 Briefly press button. 2 The following message appears in the text display: Babytherm 8010 »Configuration Mode Mode c04: Time counter photo therapy«. 3 Indication of the operating hours up to 1000 hours. 15min °C 4 Press button for 3 seconds.
Page 41
Operation Test Nurse call 1 Briefly press button. Babytherm 8010 2 The following message appears in the text display: »Configuration Mode Mode c06: Nurse Call Relay«. 3 The abbreviation for this mode appears: »Nuc«. 15min °C 4 The code for the current mode appears: »c06«. 5 Display »0«...
Page 42
Operation Read error memory 1 Press button until error mode is active. Babytherm 8010 2 The following message appears in the text display: »Configuration Mode Error memory«. 15min °C 3 The error number appears: »FXX«. < 36°C skin ° C .
Operation Oxygen therapy Only enrich the oxygen concentration under the instructions of a doctor and only in accordance with the arterially measured oxygen partial pressure in the patient’s blood. Otherwise there is a danger of hyperoxaemia (eye damage) or hypoxaemia (brain damage).
Operation Taking X-rays Babytherm 8010, Babytherm 8004 without X-ray tray Swivel radiant warmer to one side. The infant will continue to be warmed. Place the X-ray film cartridge directly underneath the patient. Position the X-ray machine above the bed surface.
Care Care Do not expose device to UV radiation as a means of The Babytherm 8004/8010 infant warmer system must be disinfecting/cleaning. Cracks may be caused in the acrylic thoroughly cleaned and disinfected glass components. — after each change of patient Only use the recommended cleaning agents and disinfectants.
Troubleshooting Troubleshooting General system failures Message Cause Remedy Red LED is lit, Power failure Check that unit is plugged into mains. Audible alarm on Check that the mains is switched on. Inform internal technical department. Call DrägerService Red Inop. LED lit, Malfunction Call DrägerService Audible alarm on.
Troubleshooting System faults radiant warmer Message Cause Remedy Yellow LED flashing, measured value The measured skin temperature deviates If the measured value is less than the set display flashing, message in text display: from the set value by more than the value: check that the skin temperature »Skin temp.
Maintenance Intervals Maintenance Intervals Always disinfect and clean the unit and accessories before any maintenance * – even when returning the unit to the supplier for repair. Always disconnect power supply before any maintenance! Use only Dräger original parts for maintenance. Gel mattress Patch over small cuts or other damage in the covering film with...
10 Skin temperature sensor connector: 19 Swivel joint for radiant warmer, ±90 top socket, yellow – for measuring skin temperature 20 X-ray tray at rear – Babytherm 8004 only lower socket, white – for measuring peripheral skin 21 Bed canopy...
What’s what Controller Babytherm 8010 °C °C Check PostScript-Fehler (invalidfont, findfont) > 38°C Inop. < 36°C skin °C mattress °C manual skin man. 1 ON/OFF indicator: 8 Text display: provides advisory messages for the user and The green LED is lit when the unit is ON. prompts the user to confirm/acknowledge.
Page 52
What’s what Display and keypad for mattress heater Babytherm 8010 1 Display of measured (actual) mattress temperature 2 Indicator for upper extended range >38 °C 3 Indicator for lower extended range <36 4 Keys (buttons) for setting the desired mattress temperature >...
What’s what Controller, rear view 1 Inputs for connecting the Nurse call from up to three equipment units at the workstation (connections for central alarm) 2 Output for connection to internal paging systems, Nurse call (connection for central alarm) 3 RS 232 interface (BabyLink), optional 4 RS 232 interface (modem), optional 5 Controller ON/OFF switch;...
Technical Data Technical Data Ambient conditions For operation Temperature C to 35 Atmospheric pressure 900 hPa to 1060 hPa rel. humidity 0 to 75 %, no condensation Air velocity in air-conditioned rooms max. 0.3 m/s For storage/transport Temperature –20 C to 60 Atmospheric pressure 700 hPa to 1060 hPa rel.
Page 55
Technical Data Phototherapy (optional) Lights Halogen 12 V / 50 W (set of 6) Opening angle Radiation intensity at a distance of 80 cm 9.5 W/m and useful area of 400 x 200 mm Irradiated useful area on the bed (mm) Relative spectral radiation intensity as a function of the wavelength Relative spectral radiation intensity...
Page 56
15-610 (Universal Medical Device Nomenclature System) Electromagnetic compatibility (EMC) according to EN 60601-1-2 Interface RS 232 Level measured according to DIN 66020 Babytherm 8004/8010 Monitor Monitor cable (Part No. 83 06 488) Wiring layout diagram 9-pin sub-D 9-pin sub-D socket...
Page 57
Input Connection of the Nurse caller from up to 3 devices (the red central alarm LED lights up for every alarm forwarded from these devices to the Babytherm 8004/8010). Potential-free make contact Interfaces, optional All interface signals are electrically isolated from the patient sector.
Construction and Description Construction and Description Babytherm consists of: Heated gel mattress — a chassis with infant bed and stand The heated gel mattress consists of the following components: — a control unit with operating panel gel mattress, aluminium heating plate and an electronic —...
Construction and Description Alarm description The Babytherm unit distinguishes three alarm levels: Warning (high potential risk) A continuous audible alarm that cannot be muted is emitted for: — device malfunction — power failure A sequence of alarms tones that can be muted, combined with a flashing red LED, is triggered for: —...
Shelf bracket mode, 100 V / 120 V / 230 V Centre distance: 37 mm G 13 822 Babytherm 8004 2M 30 404 Centre distance: 170 / 240 mm G 13 821 with baby control mode and optional X-ray...
Order List Name and Description Part No. Name and Description Part No. Oxygen therapy with hood For bronchial aspiration hood 2M 19 250 Aspiration hose, 1.5 m M 25 780 flowmeter, 0 to 16 L/min 2M 85 501 Secretion inspection window, set of 5 M 07 582 Nebuliser 2M 85 834...
Index Index O2 distributor ........12 Alarm O2 flowmeter .