Push-to-Connect Fitting
Tubing Installation
Figure
Push-to-Connect
Fitting Tubing Removal
TUBE REMOVAL TOOL
Thermo Scientific
These fittings are found on the Feed Water Inlet, Optional Remote
Dispenser and Ultrafilter (if applicable).
1. Make sure the tubing is cut off reasonably square and that no plastic
burrs or ridges are present.
2. Mark an insertion length of 11/16" from end of tube.
3. Wet the tube end with water and insert the tube straight into the
fitting until it bottoms out on the interior shoulder and the insertion
mark is no longer visible.
COLLET
TYPICAL
FITTING
11/16"
Insertion
Length
Figure 6-8. Tubing Installation
Refer to Figures 6-8 and 6-9.
1. Using the tool provided (AYX23), push the collet toward the body
while pulling on the tubing to release the tube.
TYPICAL FITTING
COLLET
TUBING
Figure 6-9. Tubing Removal
TUBING
PLACE REMOVAL TOOL UP AGAINST
COLLET AND PUSH FORWARD WHILE
SIMULTANEOUSLY PULLING THE TUBE
BACK
PULL TUBING