:VOLTage:AC
Parameters
Comments
Example
Chapter 3
CONFigure:VOLTage:AC [<
[,<
resolution
>|MIN|MAX|DEF]]
allows you to specify the range and resolution.
Parameter Name
Parameter Type
range
<
>
resolution
<
>
•
To select a standard measurement range, specify range as the input signal's
maximum expected voltage. The multimeter then selects the correct range that
will accept the input.
•
The
or
AUTO
DEFault
option for resolution defaults the integration time to 10 PLC.
DEFault
•
The
and
MIN
MAX
range:
For range:
MIN
= 0.1V;
For resolution: See Table 3-4 on page
each range.
•
To select autorange, specify
for the range and resolution parameters. In the autorange mode, the multimeter
samples the input signal before each measurement and selects the appropriate
range.
•
To specify a
or
MIN
AUTO
or
DEFault
for range;
CONF:VOLT:AC DEF,MAX
CONF:VOLT:AC AUTO,MAX
prevents the
MIN
or
Making AC Voltage Measurements
CONF:VOLT:AC 0.54,MAX
SAMP:COUN 3
READ?
enter statement
range
>|MIN|MAX|DEF|AUTO
selects the AC-coupled RMS voltage function and
Range of Values
numeric
0.1V|1V|10V|100V|300V|
MIN|MAX|DEF|AUTO
numeric
resolution
option for the range parameter enables autorange. The
parameters select the minimum or maximum values for
MAX
= 300V.
71
AUTO
or
DEF
resolution while autoranging, you must specify
MAX
CONF:VOLT:AC DEF,MIN
or
CONF:VOLT:AC AUTO,MIN
(you cannot omit the range parameter). This
MAX
resolution from being interpreted as a range setting.
Function: AC volts; range selected: 1A;
MAX resolution: 100 µA.
Take 3 readings; source is IMMediate by
default.
Place multimeter in wait-for-trigger state and
make measurements; send readings to output
buffer.
Enter readings into computer.
Multimeter Command Reference
Default Units
|MIN|MAX|DEF
for valid resolution choices for
for range or do not specify a value
or
or
volts
volts
93