Subsystem Syntax

:IMPedance:AUTO

Parameters
Example
Comments
:IMPedance:AUTO?
Example
100 Multimeter Command Reference
The
command enables or disables the automatic input impedance mode for DC

INPut

voltage measurements.
INPut
:IMPedance:AUTO OFF|ON
:IMPedance:AUTO?
INPut:IMPedance:AUTO <
impedance mode for DC voltage measurements. When disabled (
multimeter maintains its input impedance of 10MΩ for all DC voltage ranges. This
is useful to prevent a change in input impedance, caused by changing ranges, from
affecting the measurements.
Parameter Name
Parameter Type
<
mode
>
mode
(Impedance)
Range for
Impedance
Enable Automatic Input Impedance (use >10GΩ for 100mV, 1V and 10V ranges)
INP:IMP:AUTO ON
You can substitute decimal values for the
Conditions:
*RST
INPut:IMPedance:AUTO?
impedance mode is enabled or disabled: "
to the output buffer.
Query the Input Impedance Mode
INP:IMP:AUTO ON
INP:IMP:AUTO?
enter statement
mode
>
enables or disables the automatic input
Range of Values
boolean
OFF|0|ON|1
AUTO OFF
(10MΩ)
all ranges
(other ranges are at 10MΩ)
Enable automatic input impedance.
OFF
INP:IMP:AUTO OFF
returns a number to show whether the automatic input
" =
ON
1
Enable automatic input impedance.
Query multimeter to return input impedance
mode ("1").
Enter value into computer.
INPut
AUTO OFF
Default Units
None
AUTO ON (>10GΩ)
100mV, 1V and 10V
("
") and
("
") parameters.
0
ON
1
, "
" =
. The number is sent
OFF
0
Chapter 3
), the