Calibration Errors

190 HP E1312A and HP E1412A Multimeter Error Messages
621
AC rms full scale failed
622
Frequency counter failed
623
Cannot calibrate precharge
625
I/O processor does not respond
626
I/O processor failed self-test
The following errors indicate failures that may occur during a calibration.
The most common errors have descriptions here. Refer to the HP E1312A
and HP E1412A Service Manual for more information on the other errors.
701
Cal security disabled by jumper
The calibration security feature has been disabled with a jumper inside the
multimeter. When applicable, this error will occur at power-on to warn you
that the multimeter is unsecured.
702
Cal secured
The multimeter is secured against calibration.
703
Invalid secure code
An invalid calibration security code was received when attempting to
unsecure or secure the multimeter. You must use the same security code to
unsecure the multimeter as was used to secure it, and vice versa. The
security code may contain up to 12 alphanumeric characters. The first
character must be a letter.
704
Secure code too long
A security code was received which contained more than 12 characters.
705
Cal aborted
A calibration in progress is aborted when you send a device clear to the
multimeter.
706
Cal value out of range
The specified calibration value (
function and range.
707
Cal signal measurement out of range
The specified calibration value (
applied to the multimeter.
708
Cal signal frequency out of range
The input signal frequency for an ac calibration does not match the required
input frequency for calibration.
CAL:VALue
) is invalid for the present
CAL:VALue
) does not match the signal
Appendix B