[:VOLTage[:DC]]:RATio

Note
Parameters
Comments
Chapter 3
CONFigure[:VOLTage[:DC]]:RATio [<
[,<
resolution
>|MIN|MAX|DEF]]
measurements with the specified range and resolution.
------------------------------------------------ - =
DC:DC RATIO =
dc reference voltage
The ratio is calculated from the voltage applied to the HI and LO input terminals
divided by the reference voltage applied to the "Sense" HI and LO terminals.
Autoranging is automatically selected for the reference voltage measurement on the
"Sense" HI and LO terminals. The specified range in the command applies to the
signal connected to the HI and LO input terminals.
Autorange on the "Sense" terminals is from 100mV to 10V range only. Maximum
voltage you can apply to the "Sense" terminals is 10V.
Parameter Name
Parameter Type
range
<
>
(HI-LO input)
resolution
<
>
To select a standard measurement range, specify range as the input signal's
maximum expected voltage. The multimeter then selects the correct range to
accept the input.
The
or
AUTO
DEFault
option for resolution defaults the integration time to 10 PLC.
DEFault
The
and
MIN
MAX
range and resolution:
For range:
MIN
= 100mV;
For resolution:
MIN
selected range.
MAX
selected range.
range
>|MIN|MAX|DEF|AUTO
configures the multimeter for dc:dc ratio
dc signal voltage
Range of Values
numeric
100mV|1V|10V|100V|300V|
MIN|MAX|DEF|AUTO
numeric
resolution|
option for the range parameter enables autorange. The
parameters select the minimum or maximum values for
MAX
= 300V.
selects the best resolution (the smallest value) for the
selects the worst resolution (the largest value) for the
Multimeter Command Reference
Hi and LO input
------------------------------------------------------ -
Sense HI and LO input
Default Units
MIN|MAX|DEF
volts
volts