Ac Voltage Adjustments; Equipment Setup - HP E1411A Service Manual

5 1/2 digit multimeter
Table of Contents

Advertisement

AC Voltage
Adjustments
NOTE

Equipment Setup

Chapter 5
This procedure adjusts the HP E1411 AC voltage measurement accuracy.
The DC Voltage adjustment MUST be performed before the AC Voltage
adjustment.
Connect the equipment as shown in Figure 5-2
1.
Figure 5-2. AC Voltage Adjustment Setup
Set the HP E1411 as follows:
2.
Reset Multimeter ....................................................*RST
Auto Zero .................................................................... ON
Power Line Cycles (PLC) .............................................. 1
Line Freq Reference (CAL:LFR ) ...... 50 Hz or 60 Hz
Adjustments 41