CURRent[:DC]:RANGe

Parameters
Comments
Example
CURRent[:DC]:RANGe?
Example
124 Multimeter Command Reference
[SENSe:]CURRent[:DC]:RANGe <
measurements.
Parameter Name
Parameter Type
range
<
>
To select a standard measurement range, specify range as the input signal's
maximum expected current. The multimeter then selects the correct range.
MIN
selects the minimum range available with the
command: 10mA.
You must select a range using
resolution.
Specifying a fixed range disables the autorange mode set by the
CURR:DC:RANG:AUTO
The
CURR:DC:RANG
command on the same function.
CONFigure
Condition:
*RST
Set the DC Current Range to 3A
CURR:DC:RANG 3
[SENSe:]CURRent[:DC]:RANGe? [MIN|MAX]
numbers to the output buffer:
The present current range selected if
ranges available with the
selects the 100mA range, 100mA is the range returned.
CONFigure
The minimum current range available (10mA) if
The maximum current range available (3A) if
Query the DC Current Measurement Range
CURR:DC:RANG 3
CURR:DC:RANG?
enter statement
range
selects the range for DC current
>
Range of Values
numeric
0.01A|0.1A|1A|3A|MIN|MAX
selects the maximum range available: 3A.
MAX
CURRent:DC:RANGe
command.
command overrides the range setting from a previous
CURR:DC:RANG 1
DC Current range is 3A.
MIN
or
command are returned. For example, if
RANGe
Select 3A range.
Query multimeter to return the present range.
Enter value into computer.
Default Units
CURRent:DC:RANGe
before specifying
returns one of the following
MAX
is not specified. Only the
is specified.
MIN
is specified.
MAX
amps
Chapter 3

Advertisement

Table of Contents