:FRESistance?

Parameters
Comments
Example
Chapter 3
MEASure:FRESistance? [<
[,<
resolution
>|MIN|MAX|DEF]]
specify the measurement range and resolution.
Parameter Name
Parameter Type
range
<
>
resolution
<
>
To select a standard measurement range, specify range as the input signal's
maximum expected resistance. The multimeter then selects the correct range
that will accept the input.
The
or
AUTO
DEFault
The
DEF
option for the resolution parameter defaults the integration time to
10 PLC.
The
MIN
and
MAX
range and resolution:
For range:
= 100Ω;
MIN
For resolution:
MIN
selected range.
MAX
selected range.
To select autorange, specify
for the range and resolution parameters. In the autorange mode, the multimeter
samples the input signal before each measurement and selects the appropriate
range.
To specify a
MIN
or
MEAS:FRES? DEF
MIN
or
MAX
resolution from being interpreted as a range setting and the
resulting command becomes
MEAS:FRES? DEF,MAX
Related Commands:
Making 4-Wire Ohms Measurements
MEAS:FRES? 1500,MAX
range
>|MIN|MAX|DEF|AUTO
selects the 4-wire ohms function and allows you to
Range of Values
numeric
100Ω|1kΩ|10kΩ|100kΩ|1MΩ
|10MΩ|100MΩ|
MIN|MAX|DEF|AUTO
numeric
resolution
option for the range parameter enables autorange.
parameters select the minimum or maximum values for
=100MΩ
MAX
selects the best resolution (the smallest value) for the
selects the worst resolution (the largest value) for the
or
AUTO
DEF
MAX
resolution while autoranging, you must specify
(you cannot omit the range parameter). This prevents the
MEAS:FRES? DEF,MIN
.
,
,
FETCh?
INITiate
READ?
Function: 4-wire ohms; range selected: 10kW;
MAX resolution: 1Ω.
Multimeter Command Reference
Default Units
|MIN|MAX|DEF
for range or do not specify a value
or
ohms
ohms