Selecting The Close-Up Lens - Konica Minolta CHROMA METER Instruction Manual

Table of Contents

Advertisement

Selecting the Close-up Lens

To measure an extremely small surface, you can use the optional close-up lens.
For the procedure to attach the close-up lens, refer to the instruction manual of the close-
up lens.
When the close-up lens is attached, it is necessary to correct the measured value by
multiplying it by the transmittance of the lens. Since the correction value varies depending
on the lens attached, you need to set the lens type on the instrument.
Setting the wrong lens type prevents proper measurement.
The following table shows the relationship between the close-up lens model and the setting
on the instrument.
No lens
Attached lens
attached
Setting
None
* Factory setting: None
Operating Procedure
1.
While the measurement screen is
displayed, press the MENU key.
The setting screen is displayed on the LCD
screen.
2.
Press the RIGHT key or LEFT key to
display the [Meas. Cond.] screen.
Close-up lens
Close-up lens
No. 153
No. 135
No. 153
No. 135
5
1
42
[Meas. Cond.] -
[Close-up Lens]
Close-up lens
Close-up lens
No. 122
No. 122
3
3, 4
2
No. 110
No. 110