Light-Receiving Element (Sensor) - Konica Minolta CHROMA METER Instruction Manual

Table of Contents

Advertisement

Light-Receiving Element (Sensor)

The CS-150/160 is a high-accuracy tristimulus luminance color meter using three sensor
outputs (XYZ = Red, green and blue) of the spectral responsivity corresponding to human
eye sensitivity (CIE1931 color matching function).
It uses this spectral responsivity to measure an object, calculates the obtained tristimulus
values (X, Y and Z) to convert them into L
excitation purity, and display the result as a measured value.
The chromaticity coordinates (x, y) using L
the following equations:
X
x =
, y =
X + Y + Z
z
y
1.0
0.5
0.0
380
430
480
530
Wavelength (nm)
CIE1931 color matching function
and spectral response of CS-150/160
The LS-150/160 is a high-accuracy luminance meter using a sensor output of the standard
spectral luminous efficiency corresponding to human eye sensitivity.
JIS C 1609-1:2006 specifies the evaluation of the spectral responsivity of a luminance
meter as follows based on the deviation "f
V (λ) when the instrument is calibrated with a standard light source with spectral
distribution of 2856K black body radiation.
Classification
General Precision Class
'
f
1
The relative spectral responsivity of the LS-150/160 is equivalent to General Class AA (JIS
C 1609-1:2006) and conforms to Class B of DIN 5032-7. The value f
following equation:
The value S'(λ)
is expressed by the following equation:
rel
x y, L
v
x y (CIE 1931 color space) can be obtained by
V
Y
where X, Y and Z are tristimulus values.
X + Y + Z
CIE1931 color
x
matching
function
Relative spectral
responsivity of
CS-150/160
580
630
680
730
780
'" from the standard spectral luminous efficiency
1
3
P(λ)
: Relative spectral distribution of standard illuminant A
A
S(λ)
: Relative spectral responsivity of the illuminance meter
rel
V(λ): Standard spectral luminous efficiency
dλ: Measurement wavelength interval (10 nm)
λ
: Lower limit of visible wavelength range
1
λ
: Upper limit of visible wavelength range
2
98
uʼvʼ, L
T
duv, dominant wavelength, and
v
v
cp
1.0
0.5
0.0
380
430
480
530
580
Wavelength (nm)
Standard spectral luminous efficiency
and spectral response of LS-150/160
General Class AA
6
' is obtained by the
1
Standard
spectral
luminous
efficiency
Relative spectral
responsivity of
LS-150/160
630
680
730
780
General Class A
9

Advertisement

Table of Contents
loading

This manual is also suitable for:

Luminance meterCs-150Cs-160Ls-150Ls-160

Table of Contents