Hardware installation
March 2023
Figure 3-11: Inline gauge
A. Sensor module
B. Process connection
Install high pressure coned and threaded connection
The transmitter comes with an autoclave connection designed for pressure applications. To
connect the transmitter to your process:
Procedure
1. Apply a process-compatible lubricant to the gland nut threads.
2. Slip the gland nut onto the tube; then thread the collar onto the tube end.
3. Apply a small amount of process-compatible lubricant to the tube cone to help
4. Tighten the gland nut to a torque of 25 ft-lb.
3.4.5 
Rosemount 305, 306, and 304 Manifolds
The Rosemount 305 Integral Manifold is available in two designs: Traditional and Coplanar.
You can mount the traditional Rosemount 305 Integral Manifold to most primary elements
with mounting adapters in the market today.
56
The collar is reverse threaded.
prevent galling and facilitate sealing. Insert the tubing into the connection and
finger tighten the bolts.
NOTICE
A weep hole has been designed into the transmitter for safety and leak detection. If
fluid begins to leak from the weep hole, isolate the process pressure, disconnect the
transmitter, and reseal until the leak is resolved.