Measurement

Measurement
Notes
Make sure to perform white calibration before measuring. For details, refer
to "White Calibration" on page 28.
• For accurate measurements, measure under the same conditions
(ambient temperature, etc.).
1
Press the [Measure] tab to open the measurement screen.
2
Set the sample surface aperture on the sample to
measure.
3
A
[When measuring remotely]
a-1
Click "Measure" on the PC application.
Beep
The measurement results will be displayed in the [Latest Meas.] tab on the PC
application.
B
[When measuring using a trigger]
b-1
Check the box for "Use Measure Button on Inst."
b-2
Press the measurement button on the instrument.
Beep
The measurement results will be displayed on the instrument and in the [Latest
Meas.] tab on the PC application.
Continuously perform measurement in the same way to measure consecutively.
To perform the most recent measurement again, click the "Undo" key.
b-1
1