Electrical Specifications, 10 Mv/A Sensitivity - AEMC MR417 User Manual

Ac/dc current probe
Hide thumbs Also See for MR417:
Table of Contents

Advertisement

Pulse Response (MR417 and MR527)
Rise time from (10 to 90) %: ≤ 11 μs
Fall time from (90 to 10) %: ≤ 11 μs
AC noise on output: ≤ 1 mV or 1 A
Delay time at 10 %: ≤ 10 μs

4.2.3 Electrical Specifications, 10 mV/A Sensitivity

Output impedance: 215 Ω
MR417
Specified
measurement
range
Intrinsic uncertainty
MR527
Specified
measurement
range
Intrinsic uncertainty
Phase error (45 to 65) Hz
MR417
Specified measurement range
Phase shift
MR527
Specified measurement range
Phase shift
20
peak-to-peak
(0.5 to 30) Aac/dc (30 to 40) Aac/dc
≤ ± (3 % R + 8 mV)
(0.5 to 40) Aac/dc
≤ ± (3 % R + 8 mV)
(1 to 100) Aac
≤± 1.5 % R
(40 to 100) Aac/dc
≤± 1.5 % R
(1 to 20) Aac
≤ -3 °
≤ -2 °
AC/DC Current Probes MR417 & MR527
(40 to 60) Adc
≤± 1.5 % R
(100 to 140) Adc
≤± 1.5 % R
(20 to 40) Aac
≤ -2.2 °

Advertisement

Table of Contents
loading

This manual is also suitable for:

Mr527