All manuals and user guides at all-guides.com MR410 AC/DC CURRENT PROBE MR520 600V CAT III - 600A 600V CAT III - 1500A 600A 1500A OUTPUT: OUTPUT: 600A: 1 mV/A 1500 A: 1 mV/A AC/DC CURRENT PROBE AC/DC CURRENT PROBE...
Page 2
All manuals and user guides at all-guides.com Statement of Compliance Chauvin Arnoux , Inc. d.b.a. AEMC Instruments ® ® certifies that this instrument has been calibrated using standards and instruments traceable to international standards. We guarantee that at the time of shipping your instrument has met its published specifications.
• Never open the back of the instrument while connected to any circuit or input. • ALWAYS connect the Model MR410 or MR520 to the display device before clamping the probe onto the sample under test. • Always inspect the instrument and lead prior to use. Replace any defective parts immediately. • NEVER use the Model MR410 or MR520 on electrical conductors rated above 600V. 1.1 International Electrical Symbols This symbol signifies that the instrument is protected by double or rein- forced insulation. Use only specified replacement parts when servicing the instrument. This symbol on the instrument indicates a WARNING and that the operator must refer to the user manual for instructions before operat- ing the instrument. In this manual, the symbol preceding instructions...
Receiving Your Shipment Upon receiving your shipment, make sure that the contents are consistent with the packing list. Notify your distributor of any missing items. If the equip- ment appears to be damaged, file a claim immediately with the carrier and notify your distributor at once, giving a detailed description of any damage. Save the damaged packing container to substantiate your claim. Ordering Information AC/DC Current Probe Model MR410 (Lead - 1mV/A - 600A max) ..........Cat. #1200.70 AC/DC Current Probe Model MR520 (Lead - 1mV/A - 1500A max) ..........Cat. #1200.71 Both models include a 9V battery and a user manual. 1.4.1 Accessories and Replacement Parts Banana plug adapter (to non-recessed plug) ...... Cat. #1017.45 Banana (female) - BNC (male) adapter ....... Cat. #2118.46...
“pry” into or “hook” onto cables (will accept 2 x 500 MCM) or small bus bars. The MR series uses Hall effect technology. The electronics and batter- ies are self-contained in the handles. The output of the AC/DC probes is 1mV/A. When taking DC measurements, a mechanical DC zero allows the user to adjust for any DC offset. There is no output filtering - True RMS with DC components is possible. The phase shift is excellent, making the MR series well suited for power and power quality applications. Model MR410 measures 400A (600A peak), 600A Model MR520 measures 1000A (1500A peak), 1500A Both the MR410 and MR520 have proportional mV output for direct read- ings on multimeters, recorders, loggers and other instruments accepting banana plugs. An optional banana plug to BNC adaptor is available as an accessory. AC/DC Current Probe Models MR410 and MR520 GlobalTestSupply www.
All manuals and user guides at all-guides.com MR410 Features Ø = 1.18" (30mm) Cable max. Ø = 0.95" (24mm) Dual Cable max. 600V CAT III - 600A 600A OUTPUT: 600A: 1 mV/A AC/DC CURRENT PROBE Model MR 410 ® I N S T R U M E N T S Figure 1 1. Jaws...
I N S T R U M E N T S Figure 2 1. Jaws 5. Two-position range selection switch 2. Conductor 6. Zero adjustment knob 3. Protective non-slip guard 7. Lead, 5 ft (1.5m) 4. Green light (on when battery voltage ≥ 6.5V) 8. Safety banana plug Ø 4mm AC/DC Current Probe Models MR410 and MR520 GlobalTestSupply www. .com Find Quality Products Online at: sales@GlobalTestSupply.com...
All manuals and user guides at all-guides.com DMM / Display Compatibility The Models MR410 and MR520 current probes are compatible with any multimeter voltmeter or other voltage measuring instrument which has the following features: • Input jack that accepts 4mm safety banana plug • Range and resolution capable of displaying 1mV of output per amp of measured current • Voltmeter accuracy of 0.3% or better to take full advantage of the probe accuracy • Input impedance of 1MΩ/100 pF or greater When the probe is making a measurement, the current-carrying conductor is not broken and remains electrically isolated from the probe output. As a result, the probe output common may be either floated (isolated) or grounded. Warning: User Safety Always use a DMM, voltmeter or other displaying device, appropriately rated for safety.
All manuals and user guides at all-guides.com CHAPTER 3 SPECIFICATIONS MR410 Specifications 3.1.1 Electrical Measurement Range: 1 to 400A (600A Peak); 1 to 600A Accuracy*: 1 to 100A: 1.5% reading ± 1A; 100 to 400A: 2.0% reading; 400 to 600A only: 2.5% reading Phase Shift: 45 to 65Hz; 10 to 200A: ≤ 2.5° 200 to 400A: ≤ 2° Overload: 1500A and 1000A continuous up to 1kHz 3.1.2 Mechanical Jaw Opening: 1.2" (31mm) Maximum Cable Diameter: One 1.18" (30mm) or two 0.95” (24mm) or two bus bars 1.2 x 0.4" (31.5 x 10mm)
*Reference conditions: 18° to 28°C , 20 to 75% RH, external magnetic field <40 A/m, no DC component, no external current carrying conductor, test sample centered, 1MΩ ≤100pF load, zero adjustment prior to measurement [DC only] DC to 65Hz, battery voltage 9V ± 0.1V. Common Specifications (MR410 & MR520) 3.3.1 Electrical Output Signal: 1mV/A Frequency Range: DC to 10kHz @ -3 dB Noise: DC to 1kHz: ≤...
Storage Temperature Range: -40° to 176°F (-40° to +80°C) Temperature Influence: ≤ 300° ppm/°K or 1%/10°K ≤ 0.2A/°K on Zero Operating Relative Humidity: 10° to 35°C: 90 ± 5% RH (without condensation) 40° to 55°C: 70 ± 5% RH (without condensation) Humidity Influence: 10 to 90% RH @ reference temperature ≤ 0.1% R Altitude: Operating: 0 to 2000m Non-operating: 0 to 12,000m 3.3.3 Safety Electrical: EN 61010-2-32, 600V CAT III, Pollution: 2 Electromagnetic Compatibility: EN 50081-1 Class B EN 50082-2 Electrostatic discharge IEC 1000-4-2 Radiated field IEC 1000-4-3 Fast transients IEC 1000-4-4 Magnetic field at 50/60 Hz IEC 1000-4-8 AC/DC Current Probe Models MR410 and MR520 GlobalTestSupply www. .com Find Quality Products Online at: sales@GlobalTestSupply.com...
A positive reading indicates current flowing in the direction of the arrow located on the side of the jaw. A negative reading indicates current flow in the opposite direction of the arrow. • Apply the conversion ratio (1mV/A) to get the value of the current. (e.g. If the multimeter reading is 250mV , the measured current is 250A AC/DC Current Probe Models MR410 and MR520 GlobalTestSupply www. .com Find Quality Products Online at: sales@GlobalTestSupply.com...
• Apply the conversion ratio (1mV/A) to get the value of the current. 4.1.3 Indicator Lights: Green LED • The Green LED indicates that the probe is on and that the battery is good. • It will not light under low battery conditions. • Replace the 9V battery if it is not lit. Tips For Getting the Best Accuracy The Models MR410 and MR520 are capable of measuring DC and fre- quency currents over a wide range. Here are some key considerations for getting the most accuracy from your display instrument: • When using the Models MR410 and MR520 with a DMM or other meter, it is important to select the range that provides the best resolution. • Make sure that probe jaw mating surfaces are free of dust and contamination.
All manuals and user guides at all-guides.com Operation Examples DC Current Measurement Example for MR410 • Conductor carrying 25A in the direction of the arrow • DMM placed in DC volts mode • DMM displays 25.0mV = 25A 25 A DC 25.0 600V CAT III - 600A 600A OUTPUT: 600A: 1 mV/A AC/DC CURRENT PROBE Model MR 410 ®...
Page 17
All manuals and user guides at all-guides.com Measuring the DC Component of an (AC+DC) Waveform for MR410 • Conductor carrying 300A + 25.0A • DMM placed in DC volts mode • DMM displays 300.0mV = 300A 300 A DC 300.0 600V CAT III - 600A 600A OUTPUT: 600A: 1 mV/A AC/DC CURRENT PROBE Model MR 410 ®...
Page 18
600V CAT III - 1500A 1500A OUTPUT: 1500A: 1 mV/A AC/DC CURRENT PROBE Model MR 520 ® I N S T R U M E N T S Figure 5 AC/DC Current Probe Models MR410 and MR520 GlobalTestSupply www. .com Find Quality Products Online at: sales@GlobalTestSupply.com...
All manuals and user guides at all-guides.com Typical Response Curves Model MR410 Typical Accuracy in DC 1000 Primary current in amperes Model MR520 Typical Accuracy in DC 1000 10,000 Primary current in amperes AC/DC Current Probe Models MR410 and MR520 GlobalTestSupply www. .com Find Quality Products Online at:...
• Completely disconnect the probe from the circuit under test, and from the DMM or measuring instrument. • Turn the probe OFF. • Unscrew the battery compartment screw and remove cover. • Replace the battery and put the cover back on. Do not replace the battery while the probe is in use. 5.1.2 Cleaning • Clean the body of the clamp with a cloth lightly moistened with soapy water. • Wipe clean with a cloth moistened with clean water and dry. • Do not use solvent. AC/DC Current Probe Models MR410 and MR520 GlobalTestSupply www. .com Find Quality Products Online at: sales@GlobalTestSupply.com...
Page 21
All manuals and user guides at all-guides.com GlobalTestSupply www. .com Find Quality Products Online at: sales@GlobalTestSupply.com...
Instruments, not by the distributor from ® whom it was purchased. This warranty is void if the unit has been tampered with, abused or if the defect is related to service not performed by AEMC ® Instruments. Full warranty coverage and product registration is available on our website at www.aemc.com/warranty.html.
Page 23
All manuals and user guides at all-guides.com GlobalTestSupply www. .com Find Quality Products Online at: sales@GlobalTestSupply.com...
Page 24
All manuals and user guides at all-guides.com 10/15 99-MAN 100108 v11 Chauvin Arnoux , Inc. d.b.a. AEMC Instruments ® ® GlobalTestSupply www. .com Find Quality Products Online at: sales@GlobalTestSupply.com...