Specifications; Electrical Specifications - AEMC 566 User Manual

Trms leakage current clamp-on meter
Table of Contents

Advertisement

Reference Conditions: Accuracy given at 73°F ± 9°F (23°C ± 5°F),
≤ 80% RH, Conductor Centered in A, Sine wave 48-65Hz, No AC
Magnetic Field, External Magnetic Field <40A/m, True RMS (no DC
component) for V, A and mA accuracy are specified from 5% to 100%
of range. Accuracy add ±1% of Reading on Crest Factor 1.4<CF<3 at
full-scale & CF<6 at mid-scale.
3.1

Electrical Specifications

ACmA Current
Range
Res
6mA
0.001mA
60mA
0.01mA
600mA
0.1mA
Max Voltage to Ground: 600V
Zero correction: Currents smaller than approximately 0.006mA
display as zero.
Low Pass Filter
Range
6mA
0.001mA
60mA
0.01mA
600mA
0.1mA
AC Current
Range
Res
6A
1mA
60A
10mA
Max Voltage to Ground: 600V
8
CHAPTER 3

SPECIFICATIONS

(TRMS, Auto-ranging)
1.0% Rdg ± 8cts (50 to 60Hz)
2.0% Rdg ± 8cts (60 to 500Hz)
1.0% Rdg ± 5cts (50 to 60Hz)
2.0% Rdg ± 5cts (60 to 500Hz)
Res
2.0% Rdg ± 8cts
2.0% Rdg ± 5cts
(TRMS, Auto-ranging)
1.0% Rdg ± 5cts (50 to 60Hz)
2.0% Rdg ± 5cts (60 to 500Hz)
Accuracy
Accuracy
Accuracy
Clamp-on Meter Model 566
Max
660mArms
Max
66Arms