Calibration Preparation And Instrument Setup - Advanced Energy 1316 User Manual

Precision calorimeter
Table of Contents

Advertisement

9. On the 12 kW Chiller display, touch the PUMP ONLY icon and hold for 2-3 seconds
to circulate coolant.
10. Allow sufficient time for the coolant to circulate through the system before proceeding
to the next step.
NOTE: If the coolant level drops and triggers a low coolant alarm, add additional
coolant mixture as necessary to the 12 kW Chiller reservoir until the alarm turns off.
Clear the alarm on the 12 kW Chiller display.
11. On the 12 kW Chiller display, touch the PUMP ONLY icon and hold for 2-3 seconds
to stop flow circulation coolant.
12. Power OFF the 12 kW Chiller.

Calibration Preparation and Instrument Setup

1. Install the HPC-CAL calibration software on the system workstation.
2. Power ON the 12 kW Chiller.
3. On the 12 kW Chiller display, touch RUN and hold for 2-3 seconds.
4. Power ON the Calorimeter. Once the Calorimeter completes the boot process:
a. On the Calorimeter front panel, press the SETUP softkey.
b. Navigate to Instrument using the arrow keys and press ENTER.
c. Navigate to PUMP and press ENTER.
d. Press ENTER again to begin editing the parameter value.
e. Navigate to USE PUMP and press ENTER.
f. Press the UP arrow key to change the parameter value to TRUE and press
ENTER.
5. Verify that the green FLOW OK indicator is lit.
6. Power ON the 1 kW Chiller.
7. Connect cables CA-70-5 and CA-71-5 to the Power Analyzer as shown in Figure 5
below.
Figure 5: CA-70-5 and CA-71-5 connections to Power Analyzer
1316-900 Revision AA
10kW Precision Calorimeter
15