14. In the Calorimeter Check group, verify the Calorimeter Flow: and Flow Meter: fields
indicate 4.05 GPM ±0.03, and that Coolant Temp: indicates 22.1 °C ±0.4 (see Figure
11, Step 1 below).
15. (Optional) In the Step 2 group, click to select the Check power to standard check box
(Figure 11, Step 2).This setting instructs the HPC-CAL calibration software to verify
the connection between the Power Analyzer and AC Power Source.
16. In the Step 2 group, click the Check All Equipment button (Figure 11, Step 3).
17. Wait approximately 60 seconds for the HPC-CAL calibration software to verify
communication with the connected instruments. Verify the Power Source, Power
Standard, and Calorimeter checkboxes are checked. The Check All Equipment
button text should be black (Figure 11, Step 4).
18. In the Step 3 group, click the Start Auto Test button (Figure 11, Step 5).
1316-900 Revision AA
Figure 10: Step 1 group fields
Description
Technician ID or initials.
The calibration authority/procedure for
the test.
The report number must be 10
characters and should be unique. The
date, including minutes
(yymmddhhmm), is recommended to
ensure each report number is unique.
Company or department name, or test
station number.
Ambient temperature of the calibration
environment.
Ambient humidity in the calibration
environment.
Physical location of the
temperature/humidity sensor
Table 1: Step 1 group field definitions