Calibration - Advanced Energy 1316 User Manual

Precision calorimeter
Table of Contents

Advertisement

8. Connect cables CA-70-5 and CA-71-5 to the Voltage/Current RF Adapter as shown in
Figure 6 below.
Figure 6: CA-70-5 and CA-71-5 connections to Voltage/Current RF Adapter
9. Connect the Voltage/Current RF Adapter output HN RF connector to the Calorimeter
input HN RF connector.
10. Allow the system to stabilize for no less than two (2) hours before performing the
remaining steps of this procedure.

Calibration

1. Launch the HPC-CAL calibration software on the system workstation.
2. Click the Test Definition tab (see Figure 7 below).
3. Click the Read Test Definition button (see Figure 7).
Figure 7: HPC-CAL calibration software main window
4. In the Open File dialog box, select the Test Definition file titled HPC-
Cal_TestDefinitions_SNxxxx.XML.
5. Click Open to load the selected calibration test parameters into the HPC-CAL
calibration software as shown in Figure 8 below.
1316-900 Revision AA
10kW Precision Calorimeter
16