Advanced Energy 1316 User Manual page 30

Precision calorimeter
Table of Contents

Advertisement

Table 3: General Troubleshooting Guide
Symptom
No power displayed
Power won't settle or excessively long settling
time
Flow rate much less than nominal
AC calibrated power equals input power
RF calibrated power equals AC calibrated power
Extreme ambient temperature readings
No communication with 1316 Calorimeter
1316-900 Revision AA
Corrective Action
Check load DC resistance for possible failure
Check RF amp for possible power drifting
Make sure chiller set point and PID parameters
are correct
Check for blockage in fluid filter
Check calibration.ini file for proper values
Check reflection.ini file for proper values
Check for fan failure or inlet air blockage
Check for proper IP address
10kW Precision Calorimeter
25