Calorimeter/Chiller
Calorimeter
Fine Chiller
Course Chiller (Optional) 63" height, 33" width, 30" depth
AC Source (Optional)
AC Standard (Optional)
Coolant Flow Rate
Drift
Input Impedance
Communication
Cooling Fluid
Operating Temperature
Storage Temperature
Warranty
S11 Linear Magnitude (Typical)
Optional System Components (contact your AE-TEGAM representative for more information).
AC Source
AC Power Meter
Course Chiller-Air/Water Cooled Required to pre-cool cooling fluid before fine cool chiller.
P/C
Other
1
Average of ten (10) consecutive points taken 25 seconds apart, where the standard deviation of
those points is < 0.05 W.
2
The 1316 is intended for use as a primary standard. Measurement repeatability and uncertainties
will vary among laboratories depending upon various factors, including ambient environment,
user experience, and elapsed time since the most recent system calibration. TEGAM will provide
1316-900 Revision AA
56" height, 23" width, 30"/50" depth (together in single bay rack (/w/table))
9" height, 19" width, 30" depth (removed from rack)
11" height, 19" width, 25" depth (removed from rack)
58" height, 25" width, 34" depth
8" height, 12" width, 23" depth
4 gallons (15.2 Liters) per minute nominal
< 50% of uncertainty over 48 hours
50 Ω nominal
Ethernet
25% DOW-Therm SR-1, 75% distilled water
68 to 86 °F (20 to 30 °C)
14 to +122 °F (-10 to +50 °C)
1-year Parts & Workmanship for AE-TEGAM-manufactured components
Typical < 0.025 at f < 100 MHz, and <0.1 at f >= 100 MHz
Typical S
Linear Mag Performance
11
Required for routine calibration of the Calorimeter.
Required for routine calibration of the calorimeter.
For control of calorimeter and ancillary equipment.
Contact TEGAM for further details and options.
10kW Precision Calorimeter
6
6
6
4
4