Advanced Energy 1316 User Manual page 22

Precision calorimeter
Table of Contents

Advertisement

Figure 8: HPC-CAL calibration software with Test Definitions file loaded
6. Click the Device Definitions tab (see Figure 9, Step 1 below).
7. Click the Configure Definition drop-down box (Figure 9, Step 2).
8. Click Calorimeter1316 in the drop-down box (Figure 9, Step 3).
9. Click the Network tab (Figure 9, Step 4).
10. Verify that the IP address displayed in the IP Addr textbox matches the Calorimeter IP
address (Figure 9, Step 5).
11. Click the Check button. If communication is properly configured, the Calorimeter will
respond with the instrument model number (Figure 9, Step 6).
12. Repeat Steps 7 through 11 above for the AC Power Source and Power Analyzer. In
Step 9, click the tab that corresponds to the appropriate communication protocol for the
instrument being configured.
Figure 9: HPC-CAL calibration software Device Definitions tab
13. Complete the fields contained in the Step 1 group as shown in Figure 10. TEGAM has
defined the fields as described in Table 1 below, but the user may choose to input
additional or different information in any field. Note that the information will appear in
various HPC-CAL calibration software output files.
1316-900 Revision AA
10kW Precision Calorimeter
17