Service and troubleshooting
March 2023
Dropping of level close to tank bottom
Symptom
Measured value drops to zero level in the tank bottom region.
Figure 7-10: Symptom
A. Level
B. Time
C. Actual level
D. Reported level
Possible cause
Transmitter has locked on a strong tank bottom echo.
Recommended actions
•
Related information
Reference height
58
A
C
D
Verify the Reference Height is configured correctly.
B
Rosemount 1208C Level and Flow Transmitter
Reference Manual
00809-0200-7062