Maintenance and Troubleshooting Manual
00809-0100-3814
16. Connect to the meter using MeterLink. Open the detailed view in Meter
17. Use the Tools → Transducer Swap-out menu path to access the Transducer
18. Disconnect communications with the meter and close DUI.
19. If required, install the security wires on the transducer(s) cable and locking
3.3.1 
Manually modifying the Calibration Parameters
Use the Tools → Transducer Swap-out utility in MeterLink to easily update
parameters such as path lengths, delay times, and delta times for chord. This
is necessary anytime housings have to be replaced for a chord or if the meter has
performed a cold start. The wizard steps you through selecting which chords have
been changed, which components in the chords have been changed, as well as
recalculating new path lengths based on the information entered. This utility is
only available while connected to a meter. The length of the meter body is found
on the original calibration sheet supplied with the meter.
Rosemount 3814 Liquid Ultrasonic Flow Meter
fluids leaking from the threads, recheck all connections and resolve the
problem. Then, slowly repressurize the meter to line pressure.
Monitor and verify the meter is acquiring data, the transducers have good
signals and flow profiles for the chords are displayed.
Swap-out Wizard. The Transducer Swap-Out Wizard is a utility that allows
you to easily update parameters such as path lengths, delay times, and
delta times for chord. This is necessary anytime housings have to be
replaced for a chord. The wizard guides you through updating which chords
have been changed, which components in the chords have been changed,
as well as recalculating new path lengths based on the information entered.
This utility is only available while connected to a meter. Collect and
save a Maintenance Log and verify the meter is optimally performing. If
communicating with the meter via Modbus or HART, manually update the
parameters (see
Manually modifying the Calibration
ring (see
Figure
3-6) for each component you replaced.
This completes the removal and installation of the transducer housing.
Meter maintenance
January 2023
Parameters).