Meter maintenance
January 2023
13. Slide the locking ring onto the transducer housing in an orientation so that
14. Align the transducer cable connector with the indexing key on the back of
15. Apply power to the meter and slowly repressurize the meter to line
50
Transducer
Torque range Ft-
housing
lb (N-m)
LT-01/LT-03
20 (27)
LT-08/LT-09
LT-04/LT-05
the locking ring bolt can be installed through the slot of the locking ring
and into the threaded hole in the meter body. The recommended torque for
installing the locking ring bolt is 2.5 ft-lb (3.4 N-m).
Figure 3-10: Transducer installation
A. Transducer locking ring
the transducer and hand-tighten the coupling nut to secure the cable (see
Figure
3-10).
pressure. Check for leaks as the meter is pressurized. If you hear or see
Maintenance and Troubleshooting Manual
Transducer housing
A. Transducer housing
Emerson.com/Rosemount
00809-0100-3814