Emerson Rosemount 3814 Maintenance And Troubleshooting Manual page 31

Liquid ultrasonic flow meter
Hide thumbs Also See for Rosemount 3814:
Table of Contents

Advertisement

Maintenance and Troubleshooting Manual
00809-0100-3814
Table 2-2: Maintenance
MeterLink utility
MeterLink Tools → Edit/
Compare Configuration
menu
MeterLink Logs/Reports
Menu
Rosemount 3814 Liquid Ultrasonic Flow Meter
(continued)
Diagnostics
Meter performed a Cold
start
Power failure
Troubleshooting
January 2023
Action(s)
device number, or a
Modbus map file.
The meter configuration
has reset to default
values and the meter is
not configured correctly
to measure flow.
Unless the cold start
occurred after upgrading
firmware, you may need
to replace the CPU
Module.
If the Cold Start occurred
after a firmware
upgrade, you must
reconfigure the meter
from a previously saved
configuration file using
the Edit → Compare
Configuration screen.
For Rosemount Series
3810 Firmware v1.60
and later, the user
database must be either
imported from a saved
user database using
Meter → Manage Users
dialog box or manually
reentered using the
same dialog box. Then
clear the latched alarm
on the Status Summary
page.
If this was a known
power fail or restart
of the meter just
acknowledge this alarm
on the Status Summary
page.
If this was an unexpected
restart of the meter,
31