Emerson Rosemount 3814 Maintenance And Troubleshooting Manual page 27

Liquid ultrasonic flow meter
Hide thumbs Also See for Rosemount 3814:
Table of Contents

Advertisement

Maintenance and Troubleshooting Manual
00809-0100-3814
Table 2-2: Maintenance
MeterLink utility
Meter Monitor (Detailed)
view
Rosemount 3814 Liquid Ultrasonic Flow Meter
(continued)
Diagnostics
Flow Profile
Troubleshooting
January 2023
Action(s)
Flow profile ratios can be
viewed by clicking on the
arrows in the upper left
of the chart.
The flow profile ratios
for chords A and D and
chords B and C should be
equal lengths.
The flow velocity ratio
between chords A & D or
chords B & C should not
differ by more than 10%,
changes will be reflected
by the symmetry and
flow profile values.
If the difference persists,
the flow conditioner
(if installed) should be
checked for blockages
and if required the
upstream piping should
be cleaned.
Per path Speed of Sound
(SOS) differences should
not be more than .35%,
these can be viewed by
clicking on the arrows
in the upper left of the
chart. Negative values
are shown in blue.
The SOS difference can
be caused by buildup
on the transducer,
incorrect geometrical
and delay time values
and stratification at
lower velocities.
Gains and Signal to
Noise Ratios (SNR) are
displayed in decibels and
should be compared to
the initial values in the
maintenance logs taken
during the calibration or
initial start-up.
The average chord
signal amplitudes should
be compared to the
27