R&S NRP Series Getting Started page 40

Power sensors
Hide thumbs Also See for NRP Series:
Table of Contents

Advertisement

®
R&S
NRP Series
Setup
1
3-Path Diode Power Sensor
MHz to GHz, 100 pW to 200 mW (−70 dBm to +23 dBm)
SMART SENSOR TECHNOLOGY
Figure 5-3: Setup with the R&S Power Viewer
1 = Signal source
2 = R&S NRP power sensor
3 = Host interface connector
4 = R&S NRP‑ZKU cable
5 = USB connector
6 = Computer with installed R&S Power Viewer
1. NOTICE! Incorrectly connecting or disconnecting the power sensor can dam-
age the power sensor or lead to erroneous results. Ensure that you connect or
disconnect the power sensor as described in
DUT",
on page 13.
Connect the power sensor to the signal source.
2. Connect the cables as shown in
For a detailed description, refer to
on page 16.
Starting a measurement
For a detailed description, refer to the operating manual of the R&S Power
Viewer. The manual is installed automatically during the installation of the
R&S Power Viewer.
1. Start the R&S Power Viewer.
2. Execute zeroing.
Note: Turn off all measurement power signals before zeroing. An active mea-
surement signal during zeroing causes an error.
3. Switch on the test signal of the signal source.
4. Select a measurement.
Getting Started 1419.0170.02 ─ 16
2
3
NRP
5
4
Chapter 3.4, "Connecting to a
Figure
5-3.
Chapter 3.7.1.1, "Simple USB
Operating concepts
R&S Power Viewer
6
connection",
40