3.3

Considerations for test setup

Give particular attention to the following aspects when handling power sensors.
Handling the R&S NRP18S‑xx power sensor
► CAUTION! Hot surfaces. Under certain conditions, the maximum surface tem-
peratures of the power sensor can exceed the limits defined in the
EN 61010-1 standard, safety requirements for electrical equipment for mea-
surement, control and laboratory use.
Provide protection as follows:
a) Ensure that unintentional contact with the power sensor is impossible.
b) Wear heat-protective gloves when touching the power sensor after opera-
tion.
Handling the TVAC‑compliant power sensor
1. NOTICE! Avoid contamination.
Always wear clean protective gloves when handling the TVAC‑compliant
power sensor to protect the power sensor and its environment from contami-
nation.
2. NOTICE! Reduce outgassing to a minimum by following this bake-out proce-
dure.
Vacuum bake the TVAC‑compliant power sensor for 100 hours at 85 °C at a
pressure lower than 10
EMI impact on measurement results
Electromagnetic interference (EMI) can affect the measurement results.
Getting Started 1419.0170.02 ─ 16
-5
mbar.