3 Working conditions
Working temperature: 10°C to 30°C
Humidity: 15% to 80%.
4 Technical data
• Measuring path of the measuring system when reaching the maximum load: 10
mm
• Measurement uncertainty: ± 1 % of Max (measuring range)
• Weight: 560 g
• Size in mm: L 230 x W 60 x H 50
5 Exhibitions
Attach a suitable attachment (standard accessories see contents of case)
Track or Peak:
corresponding position.
Zero position:
Tare function:
Note:
The longevity of the spring measuring system is improved if the device is
set to peak during rest periods.
6 Adaptation to test benches
At all test stands with four M3 screws. This requires the disassembly of the rear half of
the housing.
7 Warnings
Incorrectly performed force measurements can lead to serious injury to persons and
damage to property and should therefore only be performed by trained and
experienced personnel.
In particular, it must be avoided that the purchased measuring instrument is subjected
to forces that exceed the maximum load (Max) of the instrument or that do not act
axially on the instrument via the force transducer; or if high pulse forces act on the
measuring instrument.
Avoid twisting the instrument, as this could cause the instrument to break and in any
case reduce the measuring accuracy.
FA-BA-e-2020
Switch by pressing and turning the knob simultaneously to the
In peak position by pressing the shift button once.
Zeroing of the display by turning the display accordingly
5