b. Quick Procedure
If you are in a hurry to fill the tanks, you can use the funnel provided to fill the tanks manually.
Begin the following procedures with only the DEV rack set in place.
q
Push the OFF side (™side) of the power breaker.
w
Fill the FIX tank with solution up to the full mark, begin careful not to splash any of the
solution into the DEV tank.
NOTE: Wash the funnel thoroughly after use.
e
Set the FIX-WASH rack into the tank, being careful not to splash any of the solution.
r
Fill the DEV tank with solution to just slightly overflow the tank.
t
Use the beaker provided to measure out the specified quantity(ex. 78 p) of starter.
y
Pour the starter into the DEV tank through the starter fluid inlet.
NOTE :
"Air rock" may occur after finishing this procedure.Be sure to activate the Circulation
pump to check if the solutions circulate normally.
13-14. Operation Testing
q
Press the RUN Button ON.
G The RUN Lamp will light up.
w
Wait until the READY lamp lights up.
e
Remove the main unit's top cover.
r
As soon as the processor is READY, insert a few sheets of test film through the feed tray and
check the following operations :
a. Are the rollers rotating smoothly?
b. Are the processing solutions circulating in the tanks?
c. Is the DEV temperature at the correct level when the READY lamp comes on?
d. Is the test film entering the processor smoothly?
e. Is the alarm sounding after the trailing edge of the film enters the processor?(See p. 50)
f. Are there any leaks in piping system?
INSTALLATION INSTRUCTIONS AND PRECAUTIONS
Full Marks
(in each processing tank)
SERVICE MANUAL