INSTALLATION INSTRUCTIONS AND PRECAUTIONS
w
Select an electrical outlet that is equal to or above the required current and voltage to operate
the processor and use this outlet for the SRX-101A only.
Insufficient current to the processor may cause overheating and/or smoking.
e
The power source voltage to the processor should be within a range of ±10% of the specified
requirement. The frequency (Hz) value should be within a range of ±2.5% of the specified
requirements.
r
Lay the power cord so that it is clear of foot traffic, sharp objects and other equipment.
Damaged plugs and cords may cause the processor to overheat and/or smoke.
t
To prevent overheating or smoking, do not connect the processor to table top outlets, junction
cords or junction outlets.
y
The power cord provided as standard equipment is 2.6 meters in length. If an extension is
needed, use only legally specified (product liability standard) extension cords. Sub-standard
extension cords may cause the processor to overheat or smoke.
u
To prevent electrocution accidents, properly ground the SRX-101A. To avoid the possibility of
explosions or electrocution, do not attempt to ground the processor to any of the following
objects.
Gas pipes
Lightening rods
Telephone ground wires
Any water line or spigot not permitted by your local water department.
SERVICE MANUAL