13-10.Leveling the Main Unit
q
Use a carpenter's level. Turn the adjustable legs supporting the main unit to balance the
processor vertically and horizontally.
w
If the carpenter's level is not available, fill the tanks with a solution up to the indicator mark,
and check the level of the solution at 6 different points on the leveling lines which are located
at right and left inner side of the DEV, FIX and WASH tank and adjust if necessary.
DEV Tank
FIX Tank
WASH Tank
The leveling line to
put the carpenter's
level.(3 points)
INSTALLATION INSTRUCTIONS AND PRECAUTIONS
SERVICE MANUAL
75