TROUBLE SHOOTING
PROBLEM
No.
(2)Bands across the film • Improper roller rotation.
perpendicular to the
direction of film
transport.
(3)Black streaks
(4)Pressure marks
4
Poor film developing
(Poor density)
5
Poor fixing quality
SERVICE MANUAL
41
CAUSE
• Dirty rubber rollers.
• Film slippage in the rollers. • Make sure all of the moving parts • Show the proper method
• Improper roller rotation.
• Dirty rubber rollers.
• Film slippage in the rollers. • Make sure all of the moving parts • Show the proper method
• Developing rack roller
surface is too rough.
• Developing rack roller
surface is dirty.
• Insufficient solution
replenishing.
Faulty supply pump (P3).
Faulty supply line.
• The developer is exhausted • Replace the developer solution.
due to overuse.
• Incorrect developer
temperature.
• Insufficient replenishing
Faulty supply pump (P4)
Faulty supply line
• The fixer is exhausted
due to over use.
(When processing small film
lots.)
• The fixer temperature is
too low.
RESPONSE
• Check the film transport
mechanism.
• Clean the rollers.
are set properly.
• Replace all damaged parts.
• Check the film transport
mechanism.
• Clean the rollers.
are set properly.
• Replace all damaged parts.
• Replace any faulty rollers.
• Clean the rollers.
• Replace the faulty parts.
• Correct the developer
temperature setting.
• Replace the faulty parts.
• Replace the fixer solution.
ADVICE To CUSTOMER
• Remind about the periodical
check and maintenance.
• Show the proper method
for setting the rollers.
for setting the racks using
the guide arrows.
• Remind about the periodical
check and maintenance.
• Show the proper method
for setting the rollers.
for setting the racks using
the guide arrows.
• Remind about when the
rollers should be cleaned.
• Show the proper method
for cleaning the rollers.
• Remind about processing
solution replacement
intervals.
• Remind about processing
solution replacement
intervals.
• Show the proper method
for maintaining optimum
fixer temperature.

Advertisement