4)When the actual temperature is not between 33: and 35:, proceed the Knob position
adjustment (Fig. G).
qLoosen the set screw of the knob which locates just opposite of the pointer.
Pay attention not to move the volume shaft.
wSet the pointer at the measured temperature value.
eTighten the set screw of the knob with the volume shaft not moving.
Next, make the adjustment according to the temperature conditions as Case 1 and Case 2.
Case 1 : When the actual temperature is over 35:.
1)Cooling down of the DEV solution
Turn off RUN switch to wait the actual temperature becomes lower than 34:
In order to check the temperature, make DEV solution circulating.
(Turn off the RUN switch immediately after turning it on, the DEV solution circulates 2 minutes.)
2)Recheck
After turn off the RUN switch, set the dial at 34:, and turn on the RUN switch again.
Return to the step 3) Actual temperature checking.
Case 2 : When the actual temperature is under 33:.
Make the above step 2) Recheck.
INSTALLATION PROCEDURE
Fig. G
SERVICE MANUAL