@5 A ttach the magnetic catch to the center for the Upper Light Shield.
@6 A ttach the steel plate to the film insertion table cover so that it hits the magnetic catch when the
cover is folded up. (See the diagram below)
@7 O pen(fold up) the film insertion table cover and make sure that the magnetic catch is strong enough
to hold it open while operating the SRX-101A.
@8 P ut back the top cover and make sure that the light shield is functioning properly from the darkroom
side.
Wall
Light Shield Panel
SRX-101A
INSTALLATION PROCEDURE
Magnetic
Catch
Steel Plate