Konica Minolta SRX-101A Service Manual page 83

Medical film processor
Hide thumbs Also See for SRX-101A:
Table of Contents

Advertisement

INSTALLATION INSTRUCTIONS AND PRECAUTIONS
13-13. Filling the Tanks
a. Normal Procedure
q
Remove the top cover of the main unit.
w
Close all the drainage valves
e
Open the facility's water utility valve.
r
Plug in the power cord.
G The RUN Button Lamp on the control panel will light up.
NOTE : Do not press the RUN Button.
t
Press and hold down the Replenishment Button (about
5 seconds) until the READY Lamp begins flashing.
G The tanks will begin to be supplied with processing
solution until the READY Lamp stops flashing, indicating that the tanks are full.
This process takes about 22 minutes at 50 Hz and about 18 minutes at 60 Hz.
y
Use the beaker(measuring cup) provided to measure out the specified quantity(ex. 78 p) of
starter.
u
Pour the starter into the DEV tank through the starter fluid gap shown as below.
i
Close the top cover.
Specified quantity
(ex. 78P)
SERVICE MANUAL
78
Measuring Cup
Starter Fluid
Press and hold down
about 5 seconds
Starter Fluid Inlet
RUN
READY
RUN